FD-103
Calibration of FD-103 Overview
Summary:
A step-by-step tutorial on calibrating your FD-103 gas detector.
Model:
FD-103-O2
FD-103-H2S
FD-103-CO
FD-103-CO-LOW
Calibration Frequency:
At least every 12 months.
Calibration Kit Items Required:
- Calibration Gas Regulator (C10)
- Tubing
- Calibration Gas Bottle
FD-103-O2 = Calibration Gas (N2)
FD-103-H2S = Calibration Gas (H2S)
FD-103-CO = Calibration Gas (CO)
FD-103-CO-LOW = Calibration Gas (CO)
Calibration of FD-103-H2S, FD-103-CO, FD-103-CO-LOW
![]() ![]() |
▶️ Tutorial Calibration VideoExample: FD-103-CO-LOWExample: FD-103-CO |
Note: The same procedure is for FD-103-H2S just make sure to enter the calibration value of your calibration reference gas concentration.
✅ STEP 1: ZERO Calibration to 0.0 ppm using Fresh Air.
- Power on your gas detector and allow 2 minutes for stabilization in fresh air.
- Expose the gas detector to fresh air or pure nitrogen for at least 2 minutes.
- If using a N2 gas bottle, use a flow rate of 0.5 LPM.
- The gas detector may or may not read precisely 0.0 ppm.
- Press the LEFT and RIGHT buttons together.
- The "z" icon will appear on the display and will flash.
- Press the RIGHT button until it beeps. This will confirm ZERO calibration.
- After 5 seconds, the "z" icon disappears, and the unit will read 0.0 ppm.
- Done!
- In the main display screen, the unit will now read 0.0 ppm in fresh air.
✅ STEP 2: SPAN Calibration.
- Ensure you first undertake a Zero Calibration.
- Hold down the "M" button which is the RIGHT button.
- Enter passcode: 1234 by using LEFT button to advance the number and RIGHT button to move to the next digit. Then hold down the RIGHT button until a BEEP.
- Press the "M" RIGHT button to advance through the menu options.
- Continue until the icon "c" is displayed. This means "calibration".
- To confirm this selection, hold down the RIGHT button until a BEEP.
- Then, you will notice the "c" icon flashing to indicate you are in calibration mode.
- Then, enter the calibration reference gas value of your gas bottle.
- Once entered, hold the RIGHT button until a BEEP.
- Now the detector is expecting to receive the calibration gas. Expose the gas analyzer to your calibration reference gas for 2 minutes or until the reading has stabilized. Use a flow rate of 0.5 LPM and connect the gas to the gas detector via the FD-103 calibration cap.
- When you enter gas, you will notice the ppm values increasing. When stable or after 2 minutes press and hold the RIGHT button to SAVE the calibration point.
- After 5 seconds, you will notice that the "c" icon has disappeared.
-
The detector is now span-calibrated, and the calibrated value is shown on the main screen.
✅ STEP 3: Verification
- Allow the gas detector to stabilize in fresh air for 5 minutes, confirming it reads 0 ppm in fresh air.
- Expose the analyzer to the calibration reference gas to verify the calibration and detection repeatability.
- If the gas detector reading is within +/- 2% F.S. of the calibration gas value, this is considered a PASS. For example, FD-103-CO-LOW has a full scale of 100 ppm. The unit was calibrated to a reference gas of 10 ppm. Then, a PASS range is +/-2% of 100 ppm = +/- 2 ppm. Therefore, a PASS is between 8 ppm and 12 ppm.
- Once verified, your unit is properly calibrated and ready for operation.
- Don't forget to document the calibration event in your gas analyzer log book. Place a reminder calibration date sticker on the unit for the next calibration event.
Calibration of FD-103-O2
![]() |
▶️ Tutorial Calibration VideoExample: FD-103-O2 |
✅ STEP 1: SPAN Calibration to 20.9% in Fresh Air
- Power on your oxygen gas detector and allow 2 minutes for stabilization in fresh air.
- Hold down the "M" button which is the RIGHT button.
- Enter passcode: 1234 by using LEFT button to advance the number and RIGHT button to move to the next digit. Then hold down the RIGHT button until a BEEP.
- Press the "M" RIGHT button to advance through the menu options.
- Continue until the icon "c" is displayed. This means "calibration".
- To confirm this selection, hold down the RIGHT button until a BEEP.
- Then, you will notice the "c" icon flashing to indicate you are in calibration mode and show 20.9%, the target calibration value.
- Press and hold the RIGHT button to SAVE the calibration point.
- After 5 seconds, you will notice that the "c" icon has disappeared.
-
The detector is now span-calibrated to 20.9%, and will display 20.9% in fresh air.
✅ STEP 2: ZERO Calibration to 0.0 ppm using Fresh Air.
- Once calibrated to fresh air, 20.9%, proceed next to zero calibration.
- Hold down the "M" button which is the RIGHT button.
- Enter passcode: 1234 by using LEFT button to advance the number and RIGHT button to move to the next digit. Then hold down the RIGHT button until a BEEP.
- Press the "M" RIGHT button to advance through the menu options.
- Continue until the icon "z" is displayed. This means "zero calibration".
- To confirm this selection, hold down the RIGHT button until a BEEP.
- The "z" icon will flash.
- Expose the gas detector to pure nitrogen for at least 2 minutes.
- If using a N2 gas bottle, use a flow rate of 0.5 LPM.
- The gas detector may or may not read precisely 0.0 ppm.
- Then, press the RIGHT button to save the zero calibration point until it beeps. This will confirm ZERO calibration.
- After 5 seconds, the "z" icon disappears, and the unit will read 0.0 ppm.
- Done!
✅ STEP 3: Verification
- Allow the oxygen detector to stabilize in fresh air for 5 minutes, confirming it reads 20.9%.
- Once verified, your unit is properly calibrated and ready for operation.
- Don't forget to document the calibration event in your gas detector logbook. Place a reminder calibration date sticker on the unit for the next calibration event.