FD-600-O2
Calibration of FD-600-O2
Summary:
A step-by-step tutorial on calibrating your FD-600-O2 gas analyzer.
Model:
FD-600-O2
Calibration Frequency:
At least every 12 months.
Calibration Kit Items Required:
![]() |
▶️ Tutorial Calibration Video |
✅ STEP 1: SPAN Calibration to 20.9% oxygen in fresh air.
- Power on the analyzer and allow 60 seconds for stabilization in fresh air.
- Verify the pump is operating.
- Check the reading: No calibration is needed if it's within ±0.1% of 20.9%.
- For calibration to precisely 20.9% undertake the following.
- Press MENU to take you to the MENU icon selection.
- Select "GasCheck" from the Menu Screen
- Enter passcode: 1111
- Select the "O2" icon.
- Verify the default span value shows 20.9% (left column).
- The current reading appears in the right column.
- Press the right button "SAVE" to confirm the 20.9% calibration gas concentration.
- Go back to the main display to confirm a reading of 20.9%.
- The analyzer is now calibrated to 20.9% oxygen in fresh air.
✅ STEP 2: ZERO Calibration to 0.0% oxygen using Nitrogen Gas.
- Expose the gas analyzer to 100% Nitrogen Gas for at least 2 minutes. The gas analyzer will approach 0.0%.
- Ensure you use a flow rate of 0.5 LPM and a T-piece as recommended.
- Press MENU to take you to the MENU icon selection.
- Select "GasZero" from the Menu Screen.
- Select the "O2" icon.
- Press the "SAVE" button to confirm the zero value.
- Go back to the main display to confirm a reading of 0.0%.
- The analyzer is now zero-calibrated.
✅ STEP 3: Verification
- Allow the oxygen analyzer to stabilize in fresh air for 5 minutes, confirming it reads 20.9%.
- Once verified, your unit is properly calibrated and ready for operation.
- Don't forget to document the calibration event in your gas analyzer log book. Place a reminder calibration date sticker on the unit for the next calibration event.