FD-90E (Yellow)
Calibration of FD-90E-YELLOW
Summary:
A step-by-step tutorial on calibrating your FD-90E-YELLOW gas leak detector.
Model:
FD-90E-YELLOW
Calibration Frequency:
At least every 12 months.
Calibration Kit Items Required:
![]() |
▶️ Tutorial Calibration Video |
✅ STEP 1: ZERO Calibration to 0 ppm.
- Power on the analyzer and wait until the countdown. Ensure this procedure occurs in fresh air or outdoor fresh air.
- Before ZERO calibration, please wait another 5 minutes to ensure the thermal equilibrium of the metal oxide semiconducting sensor.
- Press the left and right buttons at the same time to go to the main menu.
- Select the "gas zero" option, then press "save".
- Done! Then go back to the main menu by selecting "ESC".
- The main display will now be reading 0 ppm.
✅ STEP 2: SPAN Calibration to 5000 ppm of methane.
- Perform this SPAM calibration after a ZERO calibration.
- Expose the gas analyzer to the methane 5000 ppm calibration reference gas for 2 minutes or until the reading has stabilized. Use a flow rate of 0.5 LPM and the rubber calibration cap as recommended.
- The digital display ppm numbers will rise.
- Press the left and right buttons at the same time to go to the main menu.
- Select "gas calib" from the Menu Screen.
- Press "edit" then enter passcode: 1111. Then press "sure".
- Enter the span reference gas value of 5000 ppm and then press "sure".
- Press the right button to "save" the span calibration.
- Done! Then go back to the main menu by selecting "back".
- Remove the calibration cap and close the regulator gas flow.
✅ STEP 3: Verification
- Allow the gas detector to stabilize in fresh air for 5 minutes, confirming it reads 0 ppm in fresh air.
- Expose the analyzer to the calibration reference gas to verify the calibration and detection repeatability.
- If the gas detector reading is within +/- 2% F.S. of the calibration gas value, this is considered a PASS. For example, FD-90E-YELLOW has a full scale of 10,000 ppm. The unit was calibrated to a reference gas of 5000 ppm. Then, a PASS range is +/-2% of 10,000 ppm = +/- 200 ppm. Therefore, a PASS is between 4800 ppm and 5200 ppm.
- Once verified, your unit is properly calibrated and ready for operation.
- Don't forget to document the calibration event in your gas analyzer log book. Place a reminder calibration date sticker on the unit for the next calibration event.