FD-600M
Calibration of FD-600M
Summary:
A step-by-step tutorial on calibrating your FD-600M datalogger analyzer.
Model:
FD-600M (all versions except O2)
Calibration Kit Items Required:
- Calibration T-piece & Tubing
- Calibration Gas Regulator (C10)
-
Calibration gas for your target gas model (i.e. CO, H2S, EX, 4GAS, etc..)
▶️ Tutorial Calibration Video |
✅ STEP 1: ZERO Calibration to 0 ppm, 0.0 ppm or 0.00 ppm.
- Power on the analyzer and allow 120 seconds for stabilization in fresh air.
- Expose the gas analyzer to fresh air or pure nitrogen for at least 1-2 minutes.
- If using a gas bottle, use a flow rate of 0.5 LPM and a T-piece as recommended.
-
The gas analyzer may or may not read precisely 0 ppm. Depending on the model, it may or may not read 0.0 or 0.00 ppm. To zero calibrate and make it read exactly 0 ppm, 0.0 ppm, or 0.00 ppm, undertake the following steps.
- Select "GasZero" from the Menu Screen.
- Highlight the gas you would like to ZERO.
- Press the "ZERO" button to confirm the zero value.
- When complete, “CONFIG SUCCESS” is displayed.
- In the main display screen, the unit will not read 0 ppm in fresh air.
✅ STEP 2: SPAN Calibration.
- Expose the gas analyzer to your calibration reference gas for 1-2 minutes or until the reading has stabilized. Use a flow rate of 0.5 LPM and a T-piece as recommended.
- Select "GasCalib" from the Menu Screen.
- Enter passcode: 1111.
- Enter the span reference gas value for calibration you are using (left column).
- The current reading appears in the right column.
- Press the middle button "Calib" to confirm the calibration gas concentration.
- Press the middle button ">>" to save the span calibration
- Wait for the "CONFIG SUCCESS!" confirmation
✅ STEP 3: Verification
- Allow the analyzer to stabilize in fresh air for 3-4 minutes, confirming it reads 0ppm in fresh air. Expose the analyzer to the calibration reference gas to verify the calibration and detection repeatability.
- Once verified, your unit is properly calibrated and ready for operation.