FD-CO2000
Calibration of FD-CO2000
Summary:
A step-by-step tutorial on calibrating your FD-CO2000 carbon dioxide gas monitor.
Model:
FD-CO2000 (USB or AC110V)
Calibration Frequency:
At least every 12 months.
Calibration Kit Items Required: None
![]() |
▶️ Tutorial Calibration Video Example: FD-CO2000
|
✅ STEP 1: Calibration to Fresh Air (400 ppm).
- Turn on the unit and leave the CO2 monitor in outdoor fresh air for 30 minutes.
- As you are calibrating, don't forget to breath away from the unit. Your exhalation breath has high CO2 concentration and may create an error signal to the unit.
- Tap on the MENU button and make the 400 icon flash.
- Once the calibration icon is flashing, tap on the ENTER button.
- The screen will display 400ppm.
- Then hold down the ENTER button for 3 seconds until you hear a beep and then you will see a countdown timer from 250 seconds.
- Once counted down, the calibration is COMPLETE. Your unit will read about 400ppm in outdoor fresh air.
- Done!